8/25/2023 0 Comments Sirona cerec bluecam tutorial![]() If there is a number there, go ahead and remove it. Default Gateway will generally be blank.Subnet mask should be 255.255.255.0 (or it can be blank).IP scheme should be (The final two digits should be the last two numbers of the mill serial number.) Adjust if needed.Verify the MC XL IP address via Configuration | Devices | MC XL.Verify the MC XL is unreachable in the software ( Mill Phase or Configuration | Devices).A slowing blinking green light indicates that the radio does not see its partner and will definitely need to be repaired.A solid green light indicates that the radio sees its partner inside the AC unit, but they may not be "talking." Just adjusting IP addresses may fix.Have them check the cable connections and re-plug as needed. ![]() No lights indicates that the radio has become unplugged.Identify the device by describing the lights on the radio. ** Do you have a patient in the chair? If yes, see Directly Connecting - Hardwiring the AC to the Milling Unit - FAQ 11290 to get the patient taken care of quickly, and then call back to troubleshoot the wireless issue when they don't have a time sensitive procedure on hold.** CEREC - Use the information below to troubleshoot communication issues with milling units communicating via H&W radio devices.Įnvironment: Primescan and Omnicam using Hoeft & Wessel Radio